§ 882.5897. External upper limb tremor stimulator.

245 words·~1 min read·/us/cfr/t21/s§ 882.5897·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. An external upper limb tremor stimulator is a prescription device which is placed externally on the upper limb and designed to aid in tremor symptom relief of the upper limb.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Non-clinical performance testing must assess the following:
(i)Characterization of the electrical stimulation, including the following, must be performed: Waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
(ii)Impedance testing, current distribution across the electrode surface area, adhesive integrity, and shelf life testing of the electrodes and gels must be conducted.
(iii)Simulated use testing of sensor performance and the associated algorithms that determine the stimulation output must be conducted.
(2)Patient-contacting components of the device must be demonstrated to be biocompatible.
(3)Performance testing must demonstrate electrical, thermal, and mechanical safety along with electromagnetic compatibility
(EMC)of the device in the intended use environment.
(4)Software verification, validation, and hazard analysis must be performed.
(5)Physician and patient labeling must include:
(i)Summaries of electrical stimulation parameters;
(ii)Instructions on how to correctly use and maintain the device;
(iii)Instructions and explanations of all user-interface components;
(iv)Instructions on how to clean the device;
(v)A shelf life for the electrodes and gel; and
(vi)Reuse information. [83 FR 52316, Oct. 17, 2018]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 882.5897
External upper limb tremor stimulator.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.